Desarrollo y validación de un método analítico para la determinación de óxido de etileno (EO) residual y etilenclorhidrina (ECH) en dispositivos médicos (cánulas y catéteres) mediante cromatografía de gases–HeadSpace (GC-HS).

dc.contributor.advisorLeón Ávila, Emerson Eliecer
dc.contributor.authorOrtiz Cañaveral, Karen Julieth
dc.date.accessioned2023-11-22T13:29:42Z
dc.date.available2023-11-22T13:29:42Z
dc.date.issued2023
dc.description.abstractLos dispositivos médicos son usados para el diagnóstico, detección, y tratamiento de diversas enfermedades benignas y/o malignas del cuerpo humano, la esterilización de dichos dispositivos médicos es necesaria ya que busca eliminar cualquier posible invasión de microorganismos patógenos a pacientes inmunológicamente comprometidos. En este sentido, los dispositivos médicos de uso cotidiano no pueden tolerar altas temperaturas en la esterilización por calor, es por esto, que otra posibilidad de esterilización es usando el gas de Óxido de Etileno (EO), por su buena compatibilidad con los materiales, el costo relativamente bajo y la facilidad del proceso. La determinación del Óxido de Etileno (EO) residual en el material esterilizado tiene interés toxicológico dado que es un compuesto perjudicial para la salud, que absorbido por los materiales poliméricos puede llegar al paciente presentando efectos adversos como irritación, mutagenicidad y carcinogenicidad. Por lo tanto, la estimación de Óxido de Etileno (EO) y su derivado Etilenclorhidrina (ECH) residual es una de las pruebas de calidad obligatorias que se deben realizar a los dispositivos médicos esterilizados con Óxido de Etileno (EO), cuyos límites permisibles fueron establecidos por organizaciones como la ISO a través de la Norma 10993-7: 2008 "Evaluación biológica de dispositivos médicos", donde se documenta cuando un dispositivo médico es aceptable para su uso. Con el objetivo de cumplir con este análisis se propone el desarrollo de la técnica para la extracción de Óxido de Etileno residual (EO) y Etilenclorhidrina (ECH) derivado del proceso de esterilización de dos tipos de dispositivos médicos como las cánulas nasales y los catéteres, inyección por HeadSpace (HS) y la separación - determinación por Cromatografía de Gases con detector de ionización de llama. Este trabajo se llevó a cabo en el Laboratorio Fisicoquímico de TECMOL FARMACÉUTICA S.A.S. El método fue totalmente validado en términos de especificidad, idoneidad del sistema, linealidad, precisión, LD y LC, exactitud y estabilidad de las soluciones. Los resultados obtenidos en las pruebas descritas aquí demostraron un método analíticamente sólido y adecuado para los análisis de control de calidad de Óxido de Etileno (EO) y Etilenclorhidrina (ECH) en los dispositivos médicos referidos.spa
dc.description.abstractenglishMedical devices are used for the diagnosis, detection, and treatment of various benign and/or malignant diseases of the human body. The sterilization of medical devices is necessary since it seeks to eliminate any possible invasion of pathogenic microorganisms to immunologically compromised patients. In this sense, medical devices for daily use cannot tolerate high temperatures for the type of heat sterilization, which is why another possibility is the use of Ethylene Oxide (EO) gas, due to its good compatibility with materials, the relatively low cost and the ease of the process. The determination of residual Ethylene Oxide (EO) in the sterilized material is of toxicological interest since it is a compound that is harmful to health, which, absorbed by polymeric materials, can reach the patient, presenting adverse effects such as irritation, mutagenicity, and carcinogenicity. Therefore, the estimation of Ethylene Oxide residual (EO) and Ethylene Chlorohydrin (ECH) derivative is one of the mandatory quality tests that must be carried out on medical devices sterilized with Ethylene Oxide (EO), whose permissible limits were established by organizations such as ISO through Standard 10993-7: 2008 "Biological evaluation of medical devices", which documents when a medical device sterilized with Ethylene Oxide is acceptable for use. In order to comply with this analysis, the development of the technique for the extraction, of residual Ethylene Oxide (EO) and Ethylene Chlorohydrin (ECH) derived from the sterilization process of two types of single-use medical devices and disposable such as nasal cannulas and catheters, injection HeadSpace sampler (HS) and separation and quantification by Gas Chromatography with flame ionization detector (GC-FID). This experimental methodology was carried out in the Laboratory of TECMOL FARMACÉUTICA S.A.S. The method was fully validated in terms of specificity, system suitability, linearity, precision, LD and LC, accuracy and stability of solutions. The results obtained in the tests described here demonstrated an analytically sound method suitable for quality control analysis of ethylene oxide (EO) and ethylene chlorohydrin (ECH) in the medical devices referred to.spa
dc.description.degreelevelPregradospa
dc.description.degreelevelQuímico Farmacéuticospa
dc.description.sponsorshipUniversidad El Bosquespa
dc.format.mimetypeapplication/pdf
dc.identifier.instnameUniversidad El Bosquespa
dc.identifier.reponamereponame:Repositorio Institucional Universidad El Bosquespa
dc.identifier.repourlrepourl:https://repositorio.unbosque.edu.co
dc.identifier.urihttps://hdl.handle.net/20.500.12495/11478
dc.language.isospaspa
dc.publisher.facultyFacultad de Cienciasspa
dc.publisher.grantorUniversidad El Bosquespa
dc.publisher.programQuímica Farmacéuticaspa
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 Internacional*
dc.rights.accessrightsinfo:eu-repo/semantics/openAccess
dc.rights.accessrightshttps://purl.org/coar/access_right/c_abf2
dc.rights.localAcceso abiertospa
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/4.0/*
dc.subjectCánulaspa
dc.subjectCatéterspa
dc.subjectEsterilizaciónspa
dc.subjectÓxido de etilenospa
dc.subjectCromatografía de gasesspa
dc.subjectEspacio cefálicospa
dc.subject.ddc615.19
dc.subject.keywordsCannulaspa
dc.subject.keywordsCatheterspa
dc.subject.keywordsSterilizationspa
dc.subject.keywordsEthylene Oxidespa
dc.subject.keywordsGas Chromatographyspa
dc.subject.keywordsHeadSpacespa
dc.titleDesarrollo y validación de un método analítico para la determinación de óxido de etileno (EO) residual y etilenclorhidrina (ECH) en dispositivos médicos (cánulas y catéteres) mediante cromatografía de gases–HeadSpace (GC-HS).spa
dc.title.translatedDevelopment and validation of an analytical method for the determination of residual ethylene oxide (EO) and ethylenechlorohydrin (ECH) in medical devices (cannulas and catheters) using gas chromatography–HeadSpace (GC-HS).spa
dc.type.coarhttps://purl.org/coar/resource_type/c_7a1f
dc.type.coarversionhttps://purl.org/coar/version/c_ab4af688f83e57aa
dc.type.driverinfo:eu-repo/semantics/bachelorThesis
dc.type.hasversioninfo:eu-repo/semantics/acceptedVersion
dc.type.localTesis/Trabajo de grado - Monografía - Pregrado

Archivos

Bloque original
Mostrando 1 - 1 de 1
Cargando...
Miniatura
Nombre:
Trabajo de grado.pdf
Tamaño:
1.79 MB
Formato:
Adobe Portable Document Format
Descripción:
Desarrollo y validación de un método analítico para la determinación de Óxido de Etileno (EO) residual y Etilenclorhidrina (ECH) en dispositivos médicos (Cánulas y catéteres) mediante Cromatografía de Gases – HeadSpace (GC - HS)
Bloque de licencias
Mostrando 1 - 1 de 1
No hay miniatura disponible
Nombre:
license.txt
Tamaño:
1.95 KB
Formato:
Item-specific license agreed upon to submission
Descripción: